contents   index   previous   next



Linear sum (1/3 oct., scope mode)

 

The linear sum of 1/3-octave frequency bands is computed and displayed. The frequency range is determined in the 1/3 octave frequency rang settings.

 

If the measured data is a transfer function (for example measured with MLS output signal), the data are corrected for the output signal so that the value for the input only is computed. This makes it possible to plot the input level and is why it is called scope mode. Two types of scope mode is supported, either Stationary signal (power) or Transient signal (energy). The scope mode settings are found in the lower right of the dialog box displayed after clicking as shown in the figure below.

 

 

If you are measuring impulse responses, Transient signal (energy) should be selected.

 

 

Note that the correct absolute value is given only if the absolute level of the input is calibrated. In Measurement->Hardware Calibration...., click the Settings... button as shown below.

 

 

Note also that if Microphone compensation is turned on in Plot->Plot Type Settings->Frequency Response/Spectrum... as shown below,

 

 

then microphone compensation will also be used in when computing the frequency response for the Levels plot type.